Centrifuge, High Speed

Centrifuge, High Speed

Product Code : EL-C-10693

This high speed microcentrifuge spins small tubes at 200 to 15,000 rpm, up to 21,380 × g, for molecular and cell biology work. A brushless DC motor drives an autoclavable 24 × 1.5/2.0 mL aluminium rotor. Speed is set in 10 rpm steps or force in 10 × g steps, and a pulse key gives quick spin-downs.

Specifications of the high speed micro centrifuge

ComponentSpecification
Speed range200 to 15,000 rpm in 10 rpm steps; accuracy ± 20 rpm
Maximum RCF21,380 × g, set in 10 × g steps
Run time30 s to 99 min, or continuous (HOLD)
Quick spinPulse (short spin) key
DriveBrushless DC motor
Standard rotor24 × 1.5/2.0 mL, high-strength aluminium, fully autoclavable
Rotor options5 mL × 18; 0.2/0.5/1.5/2 mL × 24; 0.5 mL × 36; PCR 8-strip × 4; adapters for 0.5 and 0.2 mL tubes
Rotor designAnti-sample-overflow, to help prevent tube leaks and contamination
CoolingMulti-flow air cooling, with a temperature display
Lid and bodyLid interlock with dual interlocking safety; lid releases automatically after the rotor stops; cast casing
House codeEL-C-10693
ManufacturerLab Exports, Delhi

Lab work done with the micro centrifuge

  • Pelleting bacteria and cultured cells. Grow the culture in a laboratory incubator or on an orbital shaker. Spin 1.5 mL portions until the cells form a pellet, then pour off the medium.
  • DNA and RNA preparation. Spin down precipitated nucleic acid, wash the pellet, and run spin-column clean-ups. The result is then checked on an agarose gel run from an electrophoresis power supply.
  • Collecting droplets before a reaction. Mix the reagents on a vortex mixer, then press the pulse key. The short spin brings droplets down from the lid and walls, so the full volume sits at the bottom of the tube.
  • Writing a method in × g. RCF = 1.118 × 10-5 × r × N2, where r is the rotor radius in cm and N is the speed in rpm. A run set in × g gives the same force on a rotor of another size.

Loading and running a spin

  1. Load the tubes in balanced, opposite positions. Use the adapters for 0.5 or 0.2 mL tubes.
  2. Close the lid. The interlock stops the rotor running with the lid open.
  3. Set the speed or RCF and the run time, or choose HOLD for continuous running.
  4. Start the run, or press the short spin key for a quick pulse.
  5. When the rotor stops, the lid releases by itself and the tubes can be taken out.

Frequently Asked Questions

Is rpm the same as RCF, and is RCF the same as g?

RCF and × g are the same thing: the force on the sample as a multiple of gravity. Rpm only says how fast the rotor turns. The same rpm gives a different force on a rotor of different radius, so methods are written in × g. This unit can be set either way.

How should the tubes be balanced, and what if there is an odd number?

Put tubes holding equal volumes directly opposite each other across the centre of the rotor. With an odd number, fill one extra tube with water to the same level as the unpaired sample and place it opposite. Balance the rotor even for a pulse spin.

What is the pulse or short spin key for?

It brings the rotor quickly up to speed for a brief spin. It is used to collect droplets from the lid and walls of a tube, for example after vortex mixing, so nothing is lost when the tube is opened.

Which tubes and rotors can it take?

The standard rotor holds 24 tubes of 1.5 or 2.0 mL, and adapters take 0.5 and 0.2 mL tubes. Other rotor options are 5 mL × 18, 0.2/0.5/1.5/2 mL × 24, 0.5 mL × 36 and PCR 8-strip × 4.

Is this centrifuge refrigerated?

No refrigeration is listed for this model; it is air-cooled. Multi-flow air cooling keeps the rotor temperature low, and a display shows the temperature. For steps that must stay cold, keep the tubes on ice before and after each run.

Other laboratory centrifuges

More items are in the Centrifuges section of the Laboratory Equipment range.

To enquire, email [email protected] with the house code EL-C-10693 and the quantity, or use the Add To Cart button on this page. Tender buyers can also see the Tenders and OEM page.

Last updated: 19 November 2026

   

Office Address

Works:11/315, Lalita Park, Laxmi Nagar, Delhi, 110092

[email protected]

Follow Us